TM 5-6115-275-14
NAVFAC P-8-615-14
TO 35C2-3-452-1
(6)
Test diode CR7.
(a)
Locate diode CR7 on circuit board.
(b)
Test diode CR7 with a multimeter (set to
read ohms). Measure resistance across
CR7, reverse test leads, and measure
resistance again.
(c)
Resistance should be low (1 to 50 ohms)
one way and infinite (oo) resistance the
other way.
(d)
If low or high resistance is measured in
both directions, diode CR7 should be
replaced.
(7)
Test transistor Q 1.
(a)
Make sure that transistor Q2 is still
removed and that one lead of diode CR6
is unattached to the printed circuit board.
(b)
Locate power transistor Q on the printed
circuit board.
(c)
Test transistor with a multimeter (set to
read ohms). Refer to Figure 714 and
Table 7-2. Multimeter ( + ) and (-) refer to
the multimeter test leads.
Table 7-2. Q1 Test Chart.
Multimeter ( + )
Multimeter (-)
Meter Reading
Collector
Emitter
Infinite oo
Collector
Base
Infinite oo
Emitter
Collector
Infinite oo
Base
Collector
1 to 50 ohms
Base
Emitter
1 to 50 ohms
Emitter
Base
Infinite oo
(d)
Resistance measurements other than
those listed in Table 7-2 indicate an open
or shorted junction within Q 1, Q I should
be replaced.
(8)
Solder lead of diode CR6 to printed circuit board.
(9)
Install transistor Q2 (1, Figure 7-13) with
mica insulator between Q2 and regulator housing.
(10)
If necessary, apply a small amount of heat sink
compound (P/N 13217E3704, FSCM
97403) to transistor Q2 and housing.
Secure transistor Q2 with screws (2) and
washers (3).
(11)
Make test connections.
(a)
Connect a 30 ohm 120 watt resistor to
terminals #21 and #20 of terminal strip TB
1 (5).
(b)
Obtain a 750 ohm 2 watt rheostat.
(c)
Position rheostat so that control shaft is
facing you and the three terminal lugs are
facing up. Connect a 12-18 AWG
insulated wire to the middle terminal lug
and another wire to the left hand terminal
lug of the rheostat.
(d)
Attach the two rheostat leads to terminals
#26 and #28 of terminal strip TBI (5).
Turn the rheostat clockwise to its
maximum rotation.
(e)
Connect a 0-150 VAC, 60 HZ power
supply to terminals #24 and #27 of
terminal strip TB 1 (5).
(12)
Perform tests.
(a)
Turn on the power supply and set for 115
VAC.
(b)
Set a multimeter to read a 0-50 VDC
range.
(c)
Check for 30-35 VDC between terminals
#20 and #21 of terminal strip TB 1 (5).
Terminal #20 is positive and terminal #21
is negative.
(d)
Rotate the shaft of the test rheostat
counterclockwise to approximately eighty
percent of its maximum rotation. The
voltage between terminals #20 and #21 of
TB 1 (5) should now be zero.
(e)
Rotate the shaft of the test rheostat
clockwise while observing the voltage at
terminals #20 and #21 of TB1 (5).
(f) Voltage at terminals #20 and #21 should be between
30 and 35 volts when test rheostat is
turned to approximately eighty percent of
its clockwise rotation. If no voltage is
present, refer to step (1).
(g)
If there is a delay of two seconds or more
between decreases and increases of the
voltage at terminals #20 and #21 or if
voltage changes do not occur at the eighty
percent of maximum rotation points,
adjust R6. Refer to (h) for adjustment
procedure. If no delay is present and
voltage changes occur at the eighty
percent rotation points, testing procedure
is completed. If no voltage change
occurs, regardless of the position of the
test rheostat, refer to step (i).
7-14.2 Change 8
